MRIsafety.com

Object Info:

<- Back to list
5584

Neuroform Atlas Stent
Non-clinical testing and analysis have demonstrated that the Neuroform Atlas Stent is MR Conditional alone, or when overlapped with a second stent, and adjacent to a Stryker Neurovascular coil mass. A patient with the Neuroform Atlas Stent can be safely scanned immediately after placement of this implant, under the following conditions:

-Static magnetic field of 1.5 and 3.0 Tesla

-Maximum spatial gradient field up to 2500 Gauss/cm (25 Tesla/m)

-Maximum MR system reported whole body averaged specific absorption rate of 2 W/kg (Normal Operating Mode) and head averaged specific absorption rate of 3.2 W/kg.

Stryker Neurovascular, www.stryker.com

https://www.strykerneurovascular.com

1.5,3
Conditional 8
Coils, Filters, Stents, and Grafts
Coils, Filters, Stents, and Grafts