MRIsafety.com

Object Info:

<- Back to list
4543

Wingspan Stent System

Non-clinical testing and analysis has demonstrated that the Wingspan® Stent is MR Conditional alone, or when overlapped with a second stent. A patient with this device can be safely scanned immediately after placement of this implant, in an MR system meeting the following conditions:

  • Static magnetic field of 1.5T and 3.0T only 
  • Maximum spatial gradient magnetic field of 2,500 gauss/cm (25 T/m) or less
  • Maximum MR system reported, whole body averaged specific absorption rate (SAR) of 2 W/kg (Normal Operating Mode) and head averaged SAR of 3.2 W/kg

    Stryker Neurovascular, www.stryker.com

    https://www.strykerneurovascular.com
1.5,3
Conditional 8
Coils, Filters, Stents, and Grafts
Coils, Filters, Stents, and Grafts

  Shellock R & D Services, Inc. email: Frank.ShellockREMOVE@MRIsafety.com.
  Copyright © 2024 by Shellock R & D Services, Inc. and Frank G. Shellock, Ph.D. All rights reserved.