www.bracco.com
Home
Disclaimer
The List
Safety Information
Research summary
Screening Form
Product Testing
Ordering Books
Lectures
About Dr.Shellock
Priority E-mail

MRI Safety

www.bracco.com

Safety Information


The information on this page is limited by the terms of our disclaimer. Please Read!

Cardiac Pacemakers and ICDs

*Also, see sections on Cardiac Monitors, such as the Reveal and Confirm Products

Cardiac Pacemakers and Implantable Cardioverter Defibrillators
Cardiac pacemakers and implantable cardioverter defibrillators (ICDs) are crucial implanted devices for patients with heart conditions and serve to maintain quality of life and substantially reduce morbidity. Expanded indications for cardiac pacemakers and ICDs (e.g., heart failure, obstructive sleep apnea, and other conditions) emphasize that an increasing number of patients will be treated with these devices. Currently, these cardiovascular implants are considered a relative contraindication for patients referred for MR procedures Additionally, individuals with cardiac pacemakers and ICDs should be prevented from entering the MR environment because of potential risks. For additional information on this important topic, please refer to the following sections:

Cardiac Pacemaker: EnRhythm MRI SureScan Pacing System
Cardiac Pacemaker: Advisa DR MRI SureScan Pacing System
Cardiac Pacemaker: Ensura DR MRI SureScan Pacing System



Cardiac pacemakers and ICDs have been suggested to present potential problems to patients undergoing MR procedures from various mechanisms, including:
1) Movement of the pulse generator or lead(s);
2) Temporary or permanent modification of the function of the device;
3) Inappropriate sensing, triggering, or activation of the device;
4) Excessive heating of the leads; and
5) Induced currents in the leads.

The effects of the MR environment and MR procedures on the functional and operational aspects of cardiac pacemakers and ICDs vary, depending on several factors including the type of device, how the device is programmed, the static magnetic field strength of the MR system, and the imaging conditions used for the procedure (i.e., the anatomic region imaged, type of surface coil used, the pulse sequence, amount of radiofrequency energy used, etc.). Notably, most of the data concerning the deleterious effects of MR imaging on cardiac pacemakers involved the use of older versions (i.e., pre-1996) of these devices.

Recently, many compelling reports have been published concerning “modern-day” pacemakers (i.e. devices with decreased ferromagnetic components and more sophisticated circuitry) that indicate that certain patients may undergo MR examinations without harmful effects by following specific guidelines to minimize or prevent risks. Thus, there is growing evidence that it may be possible to perform MR procedures safely in patients with cardiovascular devices under highly controlled conditions (see below).

Implantable cardioverter defibrillators (ICDs) are medical devices designed to automatically detect and treat episodes of ventricular fibrillation, ventricular tachycardias, bradycardia, and other conditions. When a problem is identified, the device can deliver defibrillation, cardioversion, antitachycardia pacing, bradycardia pacing, or other therapy.

In general, exposure to an MR system or to an MR procedure has similar effects on an ICD as that previously described for a cardiac pacemaker, since some of the basic components are comparable. However, there are several unique aspects of ICDs that impact the possible safe performance of MR procedures in patients with these devices. Therefore, patients and individuals with these devices are generally not allowed to enter the MR environment. In addition, since ICDs also have electrodes placed in the myocardium, patients are typically not permitted to undergo MR examinations because of the inherent risks related to the presence of these conductive materials.

Similar to cardiac pacemakers, it is anticipated that, in lieu of developing a truly acceptable ICD for the MR environment, safety criteria may be determined for “modern-day” ICDs that entail using specialized programming, monitoring procedures, and MR conditions to allow patients to undergo MR examinations safely. Several recent studies have described patients with ICDs examined by MR imaging without serious problems.

However, potential problems remain for these cardiovascular devices, as indicated by recent reports by Gimbel (2009) and Mollerus et al. (2009), such that extreme caution must be exercised when scanning patients, even with “modern” cardiac pacemakers and ICDs.

Growing Evidence for Performing MR Procedures in Nonpacemaker Dependent Patients
Harmful effects related to performing MRI procedures in patients with cardiac pacemakers have been documented. Notably, in virtually every case involving a fatality, the patient apparently entered the MR environment without the staff knowing a cardiac pacemaker was present. Importantly, these deaths were poorly characterized, no electrocardiographic data were available for review, it was unknown whether these patients were pacemaker dependent, and the actual cause or mechanism of death were not confirmed. By comparison, no irreversible harm has been reported when patients with cardiac pacemakers were carefully monitored during MR procedures and/or the devices underwent reprogramming prior to the scans.

Regardless of the known hazards of subjecting a patient with a cardiac pacemaker to the MR environment, numerous patients (800+) have now undergone MR imaging during purposeful, monitored procedures performed in order to conduct necessary diagnostic examinations. These patients were safely and successfully imaged using MR systems operating at static magnetic fields ranging from 0.35- to 3-Tesla without serious adverse events.

In consideration of the above, there is increasing evidence from in vitro, laboratory, and clinical studies that strict restrictions prohibiting MR procedures in patients with “modern” cardiac pacemakers and ICDs may be modified. Similar to performing MR procedures in patients with other electronically-activated devices (e.g., bone fusion stimulators, cochlear implants, neurostimulation systems, programmable injusion pumps, etc.), scanning patients with cardiac pacemakers and ICDs involves following highly specific procedures to ensure patient safety.

Interestingly, a recent study by Gimbel (2008) reported scanning patients with cardiac pacemakers at 3-Tesla with no restrictions placed on pacemaker dependency, region scanned, device type, or manufacturer. This somewhat limited experience suggested that patients may undergo carefully tailored 3-Tesla MRI scans when pre-MRI reprogramming of the device occurs in conjunction with extensive monitoring, supervision, and follow-up.

Proposed Guidelines for Performing MRI in Nonpacemaker Dependent Patients
Guidelines have been presented by Martin et al. (2004), Roquin et al. (2004), Loewy et al. (2004), the American College of Radiology (2007) and the American College of Cardiology/American Heart Association (2007) with regard to performing MR procedures in nonpacemaker dependent patients. These guidelines include, the following:
-Establish a risk-benefit ratio for the patient
-Obtain written and verbal informed consent
-Pretest pacemaker functions using appropriate equipment outside of the MR environment
-A cardiologist/electrophysiologist should decide whether it is necessary to program the pacemaker prior to the MR examination
-A cardiologist/electrophysiologist with Advanced Cardiac Life Support (ACLS) training must be in attendance for the entire MRI examination
-The patient should be monitored continuously during the MR procedure (e.g., blood pressure, pulse rate, oxygen saturation, and ECG)
-Appropriate personnel, a crash cart, and defibrillator must be available throughout the pro¬cedure to address an adverse event
-Maintain visual and voice contact throughout the procedure with the patient
-Instruct the patient to alert the MR system operator of any unusual sensations or problems so that, if necessary, the MR system operator can immediately terminate the procedure
-After the MRI examination, a cardiologist/electrophysiologist should interrogate the pacemaker to confirm that the function is consistent with the pre-examination state

MRI at 1.5 and 3-Tesla and Cardiac Pacemakers and ICDs: Magnetic Field Interactions
As previously discussed, one important safety aspect of the MR environment on cardiac pacemakers and ICDs is related to magnetic field interactions. Component parts of pacemakers and ICDs, such as batteries, reed-switches, or transformer core materials may contain ferromagnetic materials. Therefore, substantial magnetic field interactions may exist, causing these implants to move or be uncomfortable for patients or individuals. Therefore, as an important part of evaluating pacemakers and ICDs, tests for magnetic field interactions have been conducted using MR systems operating at static magnetic field strengths ranging from 0.2-Tesla (i.e., the dedicated-extremity MR system) to 3-Tesla.

Luechinger et al. investigated magnetic field interactions for thirty-one cardiac pacemakers and thirteen ICDs in association with exposure to a 1.5-Tesla MR system (Gyroscan ACS NT, Philips Medical Systems, Best, The Netherlands). The investigators reported that “newer” cardiac pacemakers had relatively low magnetic force values compared to older devices. With regard to ICDs, with the exception of one newer model (GEM II, 7273 ICD, Medtronic, Inc., Minneapolis, MN), all ICDs showed relatively high magnetic field interactions. Luechinger et al. concluded that modern-day pacemakers present no safety risk with respect to magnetic field interactions at 1.5-Tesla, while ICDs could pose problems due to strong magnet-related mechanical forces.

The clinical use of 3-Tesla MR systems for brain, musculoskeletal, body and cardiovascular applications is increasing. Because previous investigations performed to determine MR safety for pacemakers and ICDs used MR systems with static magnetic fields of 1.5-Tesla or less, it is crucial to perform ex vivo testing at 3-Tesla to characterize magnetic field-related safety for these implants, with full appreciation that additional MR safety issues exist for these devices, as described-above.

An important aspect of determining magnetic field interactions for metallic implants involves the measurement of translational attraction. Translational attraction is assessed for metallic implants using the standardized deflection angle test recommended by the American Society for Testing and Materials (ASTM) International. According to ASTM International guidelines, the deflection angle for an implant or device is generally measured at the point of the “highest spatial gradient” for the specific MR system used for testing. Notably, the deflection angle test is commonly performed as an integral part of safety testing for metallic implants and devices.

Various types of magnets exist for commercially available 1.5- and 3-Tesla MR systems, including magnet configurations that are used for conventional “long-bore” scanners and newer “short-bore” systems. Because of physical differences in the position and magnitude of the highest spatial gradient for different magnets, measurements of deflection angles for implants using long-bore vs. short-bore MR systems can produce substantially different results for magnetic field-related translational attraction, as reported by Shellock et al.

The implications are primarily for magnetic field-related translational attraction with regard to long-bore vs. short-bore 3-Tesla MR systems (with short-bore scanners producing greater translational attraction for a given implant). Therefore, a study was conducted on fourteen different cardiac pacemakers and four ICDs to evaluate translational attraction for these devices in association with long-bore and short-bore 1.5- and 3-Tesla MR systems. Deflection angles were measured based on guidelines from the ASTM International.

In general, deflection angles for the cardiovascular implants that underwent evaluation were significantly (p<0.01) higher on 1.5- and 3-Tesla short-bore scanners compared to long-bore MR systems. For the 1.5-Tesla MR systems, three cardiac pacemakers (Cosmos, Model 283-01 Pacemaker, Intermedics, Inc., Freeport, TX; Nova Model 281-01 Pacemaker, Intermedics, Inc., Freeport, TX; Res-Q ACE, Model 101-01 Pacemaker; Intermedics, Inc., Freeport, TX) exhibited deflection angles greater than 45 degrees (i.e., exceeding the recommended ASTM International criteria) on both long-bore and short-bore 1.5-Tesla MR systems. The findings indicated that these devices are potentially problematic for patients from a magnetic field interaction consideration.

With regard to the 3-Tesla MR systems, seven implants exhibited deflection angles greater than 45 degrees on the long-bore 3-Tesla scanner, while 13 exhibited deflection angles greater than 45 degrees on the short-bore 3-T MR system (refer to The List for information on the cardiac pacemakers and ICDs that underwent testing). Importantly, the findings for magnetic field-related translational attraction were substantially different comparing the long-bore (i.e., lower values) and short-bore (i.e., higher values) MR systems.

As stated, other factors exist that may impact MR safety for these cardiac pacemakers and ICDs. Therefore, regardless of the fact that magnetic field interactions may not present a risk for some of the cardiovascular implants that have been tested, these potentially hazardous mechanisms must be considered carefully for these devices.

[MR healthcare professionals are advised to contact the respective manufacturer in order to obtain the latest safety information to ensure patient safety relative to the use of an MR procedure.]

REFERENCES
Achenbach S, Moshage W, et al. Effects of magnetic resonance imaging on cardiac pacemakers and electrodes. Am Heart J 1997;134:467-473.

Alagona P, Toole JC, et al. Nuclear magnetic resonance imaging in a patient with a DDD pacemaker. Pacing Clin Electrophysiol 1989;12:619 (letter).

Anfinsen OG, Berntsen RF, Aass H, Kongsgaard E, Amlie JP. Implantable cardioverter defibrillator dysfunction during and after magnetic resonance imaging. Pacing Clin Electrophysiol 2002;25:1400-1402.

Bailey SM, Gimbel JR, Ruggieri PM, Tchou PJ, Wilcoff BL. Magnetic resonance imaging (MRI) of the brain in pacemaker dependent patients. Pacing and Clinical Electrophysiology 2005;2:S128.

Bhachu DS, Kanal E. Implantable pulse generators (pacemakers) and electrodes: safety in the magnetic resonance imaging scanner environment. J Magn Reson Imaging 2000; 12:201-204.

Bonnet CA, Elson JJ, Fogoros RN. Accidental deactivation of the automatic implantable cardioverter defibrillator. Am Heart J 1990;3:696-697.

Calcagnini G, et al. In vitro investigation of pacemaker lead heating induced by magnetic resonance imaging: role of implant geometry. J Magn Reson Imaging. 2008;28:879-86.

Del Ojo JL, Moya F, et al. Is magnetic resonance imaging safe in cardiac pacemaker recipients? Pacing Clin Electrophysiol 2005;28:274-8.

Duru F, Luechinger R, Scheidegger MB, Luscher TF, Boesiger P, Candinas R. Pacing in magnetic resonance imaging environment: clinical and technical considerations on compatibility. Eur Heart J 2001;22:113-124.

Duru F, Luechinger R, Candinas R. MR imaging in patients with cardiac pacemakers. Radiology 2001;219:856-858.

Erlebacher JA, Cahill PT, Pannizzo F, Knowles RJR. Effect of magnetic resonance imaging on DDD pacemakers. Am J Cardio 1986;57:437-440.

Faris OP, Shein M. Food and Drug Administration perspective: Magnetic resonance imaging of pacemaker and implantable cardioverter-defibrillator patients. Circulation 2006;114:1232-1233.

Fetter J, Aram G, Holmes DR, Gray JE, Hayes DL. The effects of nuclear magnetic resonance imagers on external and implantable pulse generators. Pacing Clin Electrophysiol 1984;7:720-727.

Fiek M, Remp T, Reithmann C, Steinbeck G. Complete loss of ICD programmability after magnetic resonance imaging. Pacing Clin Electrophysiol 2004;27:1002-4.

Fontaine JM, Mohamed FB, Gottlieb C, Callans DJ, Marchlinski FE. Rapid ventricular pacing in a pacemaker patient undergoing magnetic resonance imaging. Pacing Clin Electrophysiol 1998;21:1336-1339.

Garcia-Bolao I, Albaladejo V, et al. Magnetic resonance imaging in a patient with a dual chamber pacemaker. Acta Cardiol 1998;53:33-35.

Gimbel JR. Unexpected asystole during 3T magnetic resonance imaging of a pacemaker dependent patient with a 'modern' pacemaker. Europace. 2009;11:1241-2.

Gimbel JR. Letter to the Editor. Pacing Clin Electrophysiol 2003;26:1.

Gimbel JR. Magnetic resonance imaging of implantable cardiac rhythm devices at 3-Tesla. Pacing Clin Electrophysiol. 2008;31:795-801.

Gimbel JR. Implantable pacemaker and defibrillator safety in the MR environment: new thoughts for the new millennium. RSNA Special Cross-Specialty Categorical Course in Diagnostic Radiology: Practical MR Safety Considerations for Physicians, Physicists, and Technologists 2001;69-76.

Gimbel JR, Johnson D, Levine PA, Wilkoff BL. Safe performance of magnetic resonance imaging on five patients with permanent cardiac pacemakers. Pacing Clin Electrophysiol 1996;19:913-919.

Gimbel JR, Kanal E, Schwartz KM, Wilkoff BL. Outcome of magnetic resonance imaging (MRI) in selected patients with implantable cardioverter defibrillators (ICDs). Pacing Clin Electrophysiol 2005;28:270-3.

Goldsher D, Amikam S, Boulos M, et al. Magnetic resonance imaging for patients with permanent pacemakers: initial clinical experience. Isr Med Assoc J 2006;8:91-94.

Hayes DL, Holmes DR, Gray JE. Effect of 1.5 Tesla nuclear magnetic resonance imaging scanner on implanted permanent pacemakers. J Am Coll Cardiol 1987;10:782-786.

Heatlie G, Pennell DJ. Cardiovascular magnetic resonance at 0.5-T in five patients with permanent pacemakers. J Cardiovasc Magn Reson. 2007;9:15-9.

Holmes DJ, Hayes DL, Gray JE, Merideth J. The effects of magnetic resonance imaging on implantable pulse generators. Pacing Clin Electrophysiol 1986;9:360-370.

Inbar S, Larson J, et al. Case report: nuclear magnetic resonance imaging in a patient with a pacemaker. Am J Med Sci 1993;3:174-175.

Irnich W, Irnich B, et al. Do we need pacemakers resistant to magnetic resonance imaging? Europace. 2005;7:353-65.

International Commission on Non-Ionizing Radiation Protection (ICNIRP) Statement, Medical magnetic resonance procedures: protection of patients. Health Physics 2004;87:197-216.

Juralti NM, Sparker J, Gimbel JR, et al. Strategies for the safe performance of magnetic resonance imaging in selected pacemaker patients (abstract) Circulation 2001;104 (Suppl.):3020.

Kanal E, Barkovich AJ, Bell C, et al. ACR guidance document for safe MR practices: 2007. AJR Am J Roentgenol. 2007;188:1447-1474.

Lauck G, von Smekal A, Wolke S, Seelos KC, Jung W, Manz M, et al. Effects of nuclear magnetic resonance imaging on cardiac pacemakers. Pacing Clin Electrophysiol 1995;18:1549-55.

Levine PA. Industry Viewpoint: St. Jude Medical: Pacemakers, ICDs and MRI. Pacing and Clinical Electrophysiology 2005;28:266.

Levine GN, Gomes AS, Arai AE, Bluemke DA, Flamm SD, Kanal E, Manning WJ, Martin ET, Smith JM, Wilke N, Shellock FG. Safety of magnetic resonance imaging in patients with cardiovascular devices: an American Heart Association scientific statement from the Committee on Diagnostic and Interventional Cardiac Catheterization. Circulation 2007;116:2878-2891.

Loewy J, Loewy A, Kendall EJ. Reconsideration of pacemakers and MR imaging. Radiographics 2004;24:1257-1268.

Luechinger RC. Safety Aspects of Cardiac Pacemakers in Magnetic Resonance Imaging. Swiss Institute of Technology, Zurich, Dissertation, 2002

Luechinger RC, Duru F, Zeijlemaker VA, Scheidegger MB, Boesiger P, Candinas R. Pacemaker reed-switch behavior in 0.5, 1.5, and 3.0 Tesla magnetic resonance units: Are reed switches always closed in strong magnetic fields? Pacing Clin Electrophysiol 2002;25:1419-1423.

Luechinger RC, Duru F, Scheidegger MB, Boesiger P, Candinas R. Force and torque effects of a 1.5 Tesla MRI scanner on cardiac pacemakers and ICDs. Pacing Clin Electrophysiol 2001;24:199-205.

Luechinger R, Zeijlemaker VA, Pedersen EM, Mortensen P, Falk E, Duru F, Candinas R, Boesiger P. In vivo heating of pacemaker leads during magnetic resonance imaging. Eur Heart J 2005;26:376-83.

Martin TE, Coman JA, Shellock FG, Pulling C, Fair R, Jenkins K. Magnetic resonance imaging and cardiac pacemaker safety at 1.5-Tesla. Journal of the American College of Cardiology 2004;43:1315-1324.

Mollerus M, et al. Ectopy in patients with permanent pacemakers and implantable cardioverter-defibrillators undergoing an MRI scan. Pacing Clin Electrophysiol. 2009;32:772-8.

Naehle CP, et al. Evaluation of cumulative effects of MR imaging on pacemaker systems at 1.5 Tesla. Pacing Clin Electrophysiol. 2009 Sep 30. [Epub ahead of print]

Naehle CP, Meyer C, et al. Safety of brain 3-T MR imaging with transmit-receive head coil in patients with cardiac pacemakers: pilot prospective study with 51 examinations. Radiology. 2008;249:991-1001.

Naehle CP, Strach K, et al. Magnetic resonance imaging at 1.5-T in patients with implantable cardioverter-defibrillators. J Am Coll Cardiol. 2009;54:549-55.

Nazarian S, Halperin HR. How to perform magnetic resonance imaging on patients with implantable cardiac arrhythmia devices. Heart Rhythm. 2009;6:138-43.

Nazarian S, Roguin A, et al.. Clinical utility and safety of a protocol for noncardiac and cardiac magnetic resonance imaging of patients with permanent pacemakers and implantable-cardioverter defibrillators at 1.5 Tesla. Circulation 2006;114:1277-1284.

Nordbeck P, Weiss I, Ehses P, et al. Measuring RF-induced currents inside implants: Impact of device configuration on MRI safety of cardiac pacemaker leads. Magn Reson Med. 2009;61:570-8.

Pavlicek W, Geisinger M, et al. The effects of nuclear magnetic resonance on patients with cardiac pacemakers. Radiology 1983;147:149-153.

Pulver AF, Puchalski MD, et al. Safety and imaging quality of MRI in pediatric and adult congenital heart disease patients with pacemakers. Pacing Clin Electrophysiol. 2009;32:450-6.

Roguin A, et al. Magnetic resonance imaging in individuals with cardiovascular implantable electronic devices. Europace. 2008;10:336-46.

Roguin A, Donahue JK, Bomma CS, Bluemke DA, Halperin HR. Cardiac magnetic resonance imaging in a patient with implantable cardioverter-defibrillator. Pacing Clin Electrophysiol 2005;28:336-8.

Roguin A, Zviman M, et al. Modern pacemaker and implantable cardioverter-defibrillator systems can be MRI-safe. Circulation 2004;110:475-482.

Schmiedel A, et al. Magnetic resonance imaging of the brain in patients with cardiac pacemakers. In-vitro- and in-vivo-evaluation at 1.5 Tesla. Rofo 2005;177:731-44.

Shellock FG. Guest Editorial. Comments on MRI heating tests of critical implants. Journal of Magnetic Resonance Imaging 2007;26:1182-1185.

Shellock FG. Excessive temperature increases in pacemaker leads at 3-T MR imaging with a transmit-received head coil. Radiology 2009;251:948-949.

Shellock FG. Cardiac pacemakers: growing evidence for MRI safety. Signals, No. 48, Issue 1, pp. 14-15, 2004.

Shellock FG, Magnetic Resonance Procedures: Health Effects, Safety, and Patient Management CRC Press, LLC, Boca Raton, FL, 2001.

Shellock FG, Crues JV. MR procedures: biologic effects, safety, and patient care. Radiology 2004;232:635-652.

Shellock FG, Fieno DS, Thomson TJ, Talavage TM, Berman DS. Cardiac pacemaker: in vitro assessment of MR safety at 1.5-Tesla. American Heart Journal 2006;151:436-443.

Shellock FG, Fischer L, Fieno DS. Cardiac pacemakers and implantable cardioverter defibrillators: in vitro evaluation of MRI safety at 1.5-Tesla. Journal of Cardiovascular Magnetic Resonance 2007;9:21-31.

Shellock FG, O’Neil M, Ivans V, Kelly D, O’Connor M, Toay L, Crues JV. Cardiac pacemakers and implantable cardiac defibrillators are unaffected by operation of an extremity MR system. AJR Am J Roentgenol 1999;72:165-170.

Shellock FG, Tkach JA, Ruggieri PM, Masaryk TJ. Cardiac pacemakers, ICDs, and loop recorder: Evaluation of translational attraction using conventional (“long-bore”) and “short-bore” 1.5- and 3.0-Tesla MR systems. Journal of Cardiovascular Magnetic Resonance 2003;5:387-397.

Shellock FG, Valencerina S, Fischer L. MRI-related heating of pacemaker at 1.5- and 3-Tesla: Evaluation with and without pulse generator attached to leads. Circulation 112;Supplement II:561, 2005.

Shinbane J, Colletti P, Shellock FG. MR in patients with pacemakers and ICDs: Defining the issues. Journal of Cardiovascular Magnetic Resonance 2007;9:5-13.

Sommer T, Vahlhaus C, et al. MR imaging and cardiac pacemakers: in vitro evaluation and in vivo studies in 51 patients at 0.5 T. Radiology 2000;215:869-879.

Sommer T, Naehle CP, Yang A, Zeijlemaker V, et al. Strategy for safe performance of extrathoracic magnetic resonance imaging at 1.5 Tesla in the presence of cardiac pacemakers in non-pacemaker-dependent patients: a prospective study with 115 examinations. Circulation 2006;114:1285-92

Sutton R, Kanal E, Wilkoff BL, et al. Safety of magnetic resonance imaging of patients with a new Medtronic EnRhythm MRI SureScan pacing system: clinical study design. Trials. 2008 Dec 2;9:68.

Tandri H, et al. Determinants of gradient field-induced current in a pacemaker lead system in a magnetic resonance imaging environment. Heart Rhythm. 2008;5:462-8

Vahlhaus C, Sommer T, et al. interference with cardiac pacemakers by magnetic resonance imaging: Are there irreversible changes at 0.5 Tesla? Pacing Clin Electrophysiol 2001;24(Pt. I):489-95.

Wollmann C, Grude M, et al. Safe performance of magnetic resonance imaging on a patient with an ICD. Pacing Clin Electrophysiol 2005;28:339-42.

Zimmermann BH, Faul DD. Artifacts and hazards in NMR imaging due to metal implants and cardiac pacemakers. Diagn Imaging Clin Med 1984;53:53-56.

 
(c) 2009/2010 by Shellock R & D Services, Inc. and Frank G. Shellock, Ph.D. All Rights Reserved. All copyrights and pertinent trademarks are owned by Shellock R & D Services, Inc. and Frank G. Shellock, Ph.D. No part of the MRIsafety.com web site may be reproduced, stored in a retrieval system, or transmitted in any form or by any means, physical, electronic or otherwise, without the prior written permission of Shellock R & D Services, Inc. or Frank G. Shellock, Ph. D.

Request for permission to reproduce any information contained on the MRIsafety.com web site should be addressed to: frank.shellock@gte.net

Be sure to read our disclaimer.