Bioeffects of Radiofrequency Fields
Bioeffects of Radiofrequency Fields
The majority of the radiofrequency (RF) power transmitted for MR imaging or spectroscopy (including carbon decoupling, fast spin echo pulse sequences, magnetization transfer contrast pulse sequences, etc.) procedures is transformed into heat within the patient’s tissue as a result of resistive losses. Not surprisingly, the bioeffects associated with exposure to RF radiation are related to the thermogenic aspects of this electromagnetic field.
Prior to 1985, there were no reports concerning the thermophysiologic responses of human subjects exposed to RF radiation during MR procedures. Since then, many investigations have characterized the thermal effects of MR procedure-related heating.
SPECIFIC ABSORPTION RATE
Thermoregulatory and other physiologic changes that a human subject exhibits in response to exposure to RF radiation are dependent on the amount of energy that is absorbed. The dosimetric term used to describe the absorption of RF radiation is the specific absorption rate (SAR). The SAR is the mass normalized rate at which RF power is coupled to biologic tissue and is typically indicated in units of watts per kilogram (W/kg). The relative amount of RF radiation that an individual encounters during an MR procedure is designated as the whole-body-averaged SAR. Other SAR levels relative to the body part exposed or peak SAR level (i.e., the amount in one gram of tissue) may also be reported by the MR system.
Measurements or estimates of SAR are not trivial, particularly in human subjects. There are several methods of determining this parameter for the purpose of RF energy dosimetry. The SAR that is produced during an MR procedure is a complex function of numerous variables including the frequency (i.e., determined by the strength of the static magnetic field of the MR system), the type of RF pulse used (e.g., 90° vs. 180° pulse), the repetition time, the type of transmit RF coil used, the volume of tissue contained within the transmit RF coil, the configuration of the anatomical region exposed, as well as other factors.
With regard to RF energy, the U.S. Food and Drug Administration currently indicates that MR procedures that exceed certain SAR values may pose significant risks (please refer to Appendix I to review these values).
THERMOPHYSIOLOGIC RESPONSES TO MR PROCEDURE-RELATED HEATING
Thermophysiologic responses to MR procedure-related heating depend on multiple physiologic, physical, and environmental factors. These include the duration of exposure, the rate at which energy is deposited, the response of the patient’s thermoregulatory system, the presence of an underlying health condition, and the ambient conditions within the MR system.
In regards to the thermoregulatory system, when subjected to a thermal challenge, the human body loses heat by means of convection, conduction, radiation, and evaporation. Each mechanism is responsible to a varying degree for heat dissipation, as the body attempts to maintain thermal homeostasis. If the thermoregulatory effectors are not capable of totally dissipating the heat load, an accumulation or storage of heat occurs along with an elevation in local and/or overall tissue temperatures.
Various underlying health conditions may affect an individual’s ability to tolerate a thermal challenge including cardiovascular disease, hypertension, diabetes, fever, old age, and obesity. In addition, medications including diuretics, beta-blockers, calcium blockers, amphetamines, and sedatives can alter thermoregulatory responses to a heat load. Importantly, certain medications have a synergistic effect with RF radiation with respect to tissue heating. The environmental conditions (i.e., ambient temperature, relative humidity, and airflow) that exist in the MR system will also affect tissue temperature changes associated with RF energy-induced heating.
The first study of human thermal responses to RF radiation-induced heating during an MR procedure was conducted by Schaefer et al. Temperature changes and other physiologic parameters were assessed in volunteer subjects exposed to a relatively high, whole-body-averaged SAR (approximately 4.0-W/kg). The data indicated that there were no excessive temperature elevations or other deleterious physiologic consequences related to exposure to RF radiation.
Several studies were subsequently conducted involving volunteer subjects and patients undergoing clinical MR procedures, with the intent of obtaining information that would be applicable to patients typically encountered in the MR setting. These investigations demonstrated that changes in body temperatures were relatively minor (i.e., less than 0.6 degrees C). While there was a tendency for statistically significant increases in skin temperatures to occur, these were of no serious physiologic consequences.
Various studies reported a poor correlation between body temperature and skin temperature changes versus whole-body-averaged SARs associated with clinical MR procedures. These findings are not surprising considering the range of thermophysiologic responses that are possible in human subjects relative to a given SAR level. For example, as previously indicated, an individual’s thermoregulatory system can be greatly impacted by the presence of an underlying condition or medication that can impair the ability to dissipate heat.
An extensive investigation by Shellock et al. was conducted in volunteer subjects exposed to MR procedures performed at a whole-body-averaged SAR of 6.0-W/kg. To date, this is the highest level of RF energy that human subjects have been exposed to in associaton with MR procedures. Tympanic membrane temperature, six different skin temperatures, heart rate, blood pressure, oxygen saturation, and skin blood flow were monitored. The findings indicated that an MR procedure performed at a whole body averaged SAR of 6.0-W/kg can be physiologically tolerated by an individual with normal thermoregulatory function.
VERY-HIGH FIELD MR SYSTEMS
Many MR systems now operate at a static magnetic field strength of 3-Tesla, several operate at 4-Tesla and 7-Tesla, and one operates at 9.4-Tesla. These very-high-field MR systems are capable of depositing RF power that exceed those associated with a 1.5-Tesla MR system. Therefore, investigations are needed to characterize thermal responses in human subjects to determine potential thermogenic hazards associated with the use of these powerful MR devices, especially since frequency related differences likely exist. To date, with the exception of the work conducted by Kangarlu et al. (8-Tesla) and Shrivastava, et al. (9.4-Tesla) there have been few investigations of MR procedure-related heating performed using very-high-field MR systems.
REFERENCES
Atalar E. Radiofrequency safety for interventional MRI procedures. Acad Radiol. 2005;12:1149-1157.
Boss A, Graf H, Berger A, et al. Tissue warming and regulatory responses induced by radio frequency energy deposition on a whole-body 3-Tesla magnetic resonance imager. J Magn Reson Imaging. 2007;26:1334-9.
Bottomley PA. Turning up the heat on MRI. J Am Coll Radiol. 2008;5:853
Bottomley PA, Edelstein WA. Power deposition in whole body NMR imaging. Med Phys 1981;8: 510-512.
Budinger TF. Nuclear magnetic resonance (NMR) in vitro studies: known thresholds for health effects. J Comput Assisted Tomog 1981;5:800-811.
Drinkwater BL, Horvath SM: Heat tolerance and aging. Med Sci Sport Exer 1979;11:49-55.
Gorny KR, et al. Calorimetric calibration of head coil SAR estimates displayed on a clinical MR scanner. Phys Med Biol. 2008;53:2565-76.
Gowland PA, De Wilde J. Temperature increase in the fetus due to radio frequency exposure during magnetic resonance scanning. Phys Med Biol. 2008;53:L15-8.
Hand JW, Li Y, et al. Prediction of specific absorption rate in mother and fetus associated with MRI examinations during pregnancy. Magn Reson Med. 2006;55:883-893.
Hartwig V, Giovannetti G, et al. Biological effects and safety in magnetic resonance imaging: a review. Int J Environ Res Public Health. 2009;6:1778-98.
International Commission on Non-Ionizing Radiation Protection (ICNIRP) Statement, Medical magnetic resonance procedures: protection of patients. Health Physics 2004;87:197-216.
International Electrotechnical Commission (IEC), Medical Electrical Equipment, Particular requirements for the safety of magnetic resonance equipment for medical diagnosis, International Standard IEC 60601-2-33, 2002.
Kangarlu A, Shellock FG, Chakeres D. 8.0-Tesla MR system: Temperature changes associated with radiofrequency-induced heating of a head phantom. J Magn Reson Imaging 2003;17:220-226.
Kangarlu A, Ibrahim TS, Shellock FG. Effects of coil dimensions and field polarization on RF heating inside a head phantom. Magnetic Resonance Imaging 2005;23:53-60.
Kenny WL. Physiological correlates of heat intolerance. Sports Med 1985;2:279-286.
Jauchem JR. Effects of drugs on thermal responses to microwaves. Gen Pharmacol 1985;16:307-310.
Oh S, Webb AG, Neuberger T, Park B, Collins CM. Experimental and numerical assessment of MRI-induced temperature change and SAR distributions in phantoms and in vivo. Magn Reson Med. 2009 Sep 25 [Epub ahead of print]
Rowell LB. Cardiovascular aspects of human thermoregulation. Circ Res 1983;52:367-379.
Schaefer DJ. Health effects and safety of radiofrequency power deposition associated with magnetic resonance procedures. In: Shellock FG, ed. Magnetic resonance procedures: health effects and safety. Boca Raton, FL: CRC Press, 2001;55-74.
Shellock FG. Radiofrequency energy-induced heating during MR procedures: a review. J Magn Reson Imaging 2000;12:30-36.
Shellock FG, Crues JV. MR procedures: biologic effects, safety, and patient care. Radiology 2004;232:635-652.
Shellock FG, Crues JV. Temperature, heart rate, and blood pressure changes associated with clinical MR imaging at 1.5-T. Radiology 1987;163:259-262.
Shellock FG, Crues JV. Corneal temperature changes associated with high-field MR imaging using a head coil. Radiology 1988;167:809-811.
Shellock FG, Crues JV. Temperature changes caused by clinical MR imaging of the brain at 1.5 Tesla using a head coil. Am J Neuroradiol 1988;9:287-291.
Shellock FG, Rothman B, Sarti D. Heating of the scrotum by high-field-strength MR imaging. Am J Roentgenol 1990;154:1229-1232.
Shellock FG, Schatz CJ. Increases in corneal temperature caused by MR imaging of the eye with a dedicated local coil. Radiology 192;185:697-699.
Shellock FG, Schaefer DJ. Radiofrequency energy-induced heating during magnetic resonance procedures: laboratory and clinical experiences In: Shellock FG, ed. Magnetic resonance procedures: health effects and safety. Boca Raton, FL: CRC Press, 2001;75-96.
Shellock FG, Schaefer DJ, Crues JV. Alterations in body and skin temperatures caused by MR imaging: Is the recommended exposure for radiofrequency radiation too conservative? Br J Radiol 1989; 62:904-909.
Shellock FG, Schaefer DJ, Kanal E. Physiologic responses to MR imaging performed at an SAR level of 6.0 W/kg. Radiology 1994;192:865-868.
Shrivastava D, Hanson T, et al. Radiofrequency heating at 9.4T: in vivo temperature measurement results in swine. Magn Reson Med. 2008;59:73-8.
Shuman WP, et al. Superficial and deep-tissue increases in anesthetized dogs during exposure to high specific absorption rates in a 1.5-T MR imager. Radiology 1988;167:551-554.
Stuchly MA, Abrishamkar H, Strydom ML. Numerical Evaluation of Radio Frequency Power Deposition in Human Models during MRI. Conf Proc IEEE Eng Med Biol Soc. 2006;1:272-275.
U.S. Department of Health and Human Services, FDA, CDRH, Guidance for Industry and FDA Staff. Criteria for Significant Risk Investigations of Magnetic Resonance Diagnostic Devices, July 14, 2003.
Wang Z, Lin JC, Vaughan JT, Collins CM. Consideration of physiological response in numerical models of temperature during MRI of the human head.J Magn Reson Imaging. 2008;28:1303-1308.
Weintraub MI, Khoury A, Cole SP. Biologic effects of 3 Tesla (T) MR imaging comparing traditional 1.5-T and 0.6-T in 1,023 consecutive outpatients. J Neuroimaging. 2007;17:241-5.