MRIsafety.com

Object Info:

<- Back to list
6719

Neuroform EZ Stent
Non-clinical testing and analysis have demonstrated that
Neuroform Stent is MR Conditional alone, or when
overlapped with a second stent, and adjacent to a Stryker
Neurovascular coil mass. A patient with Neuroform Stent
can be safely scanned immediately after placement of this
implant, under the following conditions:
• static magnetic field of 1.5 and 3.0 Tesla
• spatial gradient field up to 2500 Gauss/cm (25 Tesla/m)
• Normal Operational Mode for gradients and SAR (maximum
whole body averaged specific absorption rate (SAR) of
lower than 2.0 W/kg and maximum head SAR of lower
than 3.2 W/kg) for a total active MR scan time (with RF
exposure) of 15 minutes or less per scan sequence.

Stryker Neurovascular, www.stryker.com

https://www.strykerneurovascular.com

1.5,3
Conditional 8
Coils, Filters, Stents, and Grafts
Coils, Filters, Stents, and Grafts


  Shellock R & D Services, Inc. email: Frank.ShellockREMOVE@MRIsafety.com.
  © 2004- by Shellock R & D Services, Inc. and Frank G. Shellock, Ph.D. All rights reserved. (v3.1.109)